Measuring phosphate with the colorimeter

Phosphate standard dilutions

«  Vanadate-molybdate method   ::   Contents   ::   Calibration Data Downloads  »

Phosphate standard dilutions

Below are methods for preparing 5 mL phosphate standards in three different concentration ranges. Adjust volumes as necessary for preparing larger volumes (e.g. 10 mL or 100 mL of standards).

  1. Low range: 0 - 5 ppm e.g. Ascorbic acid, Hach PhosVer3, API test kit
  2. Mid-range: 0 - 30 ppm e.g. Ascorbic acid method, API test kit
  3. High range: 0 - 90 ppm e.g.Vanadate-Molybdate Method

In each case, the first step is preparing a working stock by diluting a commercially available phosphate standard. This working stock is then diluted further resulting in a set of 8 phosphate concentrations (plus a distilled water blank). In addition to the equipment listed on the appropriate phosphate assay page, you will also need the following equipment and reagents if you plan to collect phosphate calibration data:

Low range (0.00 - 5.00 ppm PO4)

  • Prepare a 5 ppm PO4 working stock: 0.50 mL of the 1,000 ppm standard in 100 mL distilled water in volumetric flask. Fill to the line with distilled water.
Tube PO4 concentration (ppm) Volume of 5 ppm working stock (mL) Volume of distilled water (mL)
1 0.000 0.000 5.00
2 0.625 0.625 4.375
3 1.250 1.250 3.75
4 1.875 1.875 3.125
5 2.500 2.500 2.50
6 3.125 3.125 1.875
7 3.750 3.750 1.25
8 4.375 4.375 0.625
9 5.000 5.000 0.00

Mid range (0.00 - 30 ppm PO4)

  • Prepare a 30 ppm PO4 working stock: 3 mL of the 1,000 ppm standard in 100 mL distilled water in volumetric flask. Fill to the line with distilled water.
Tube PO4 concentration (ppm) Volume of 30 ppm working stock (mL) Volume of distilled water (mL)
1 0.00 0.000 5.00
2 3.75 0.625 4.375
3 7.50 1.250 3.75
4 11.25 1.875 3.125
5 15.00 2.500 2.50
6 18.75 3.125 1.875
7 22.50 3.750 1.25
8 26.25 4.375 0.625
9 30.00 5.000 0.00

High range (0.00 - 90 ppm PO4)

  • Prepare a 90 ppm PO4 working stock: 9 mL of the 1,000 ppm standard in 100 mL distilled water in volumetric flask. Fill to the line with distilled water.
Tube PO4 concentration (ppm) Volume of 90 ppm working stock (mL) Volume of distilled water (mL)
1 0.00 0.000 5.00
2 11.25 0.625 4.375
3 22.50 1.250 3.75
4 33.75 1.875 3.125
5 45.00 2.500 2.50
6 56.25 3.125 1.875
7 67.50 3.750 1.25
8 78.75 4.375 0.625
9 90.00 5.000 0.00

«  Vanadate-molybdate method   ::   Contents   ::   Calibration Data Downloads  »